Z-0963-2020 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-0963-2020 was initiated by Cook Inc. on January 4, 2019 and is designated Class II. Reason for recall: The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip. The recall status is terminated (terminated May 29, 2020). Affected quantity: 13388 total.

Recall Details

Product Type
Devices
Report Date
February 12, 2020
Initiation Date
January 4, 2019
Termination Date
May 29, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13388 total

Product Description

Roadrunner Hydrophilic PC Wire Guide, Guidewire, Global Product Number G18629, G17866

Reason for Recall

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Distribution Pattern

Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.

Code Information

7853231 7853232 7853233 7853239 7853240 7853241 7853242 7853243 7853245 7853247 7901140 7901141 7982113 7982118 7982133 8050291 8241613 8241616 8252179 8252181 8252186 8258553 8283322 8407934 8460985 8474895 8474898 8474905 8474906 8474908 8474909 8474910 7936207