Z-0966-2020 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-0966-2020 was initiated by Cook Inc. on January 4, 2019 and is designated Class II. Reason for recall: The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip. The recall status is terminated (terminated May 29, 2020). Affected quantity: 13388 total.

Recall Details

Product Type
Devices
Report Date
February 12, 2020
Initiation Date
January 4, 2019
Termination Date
May 29, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13388 total

Product Description

Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15302 G14326

Reason for Recall

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Distribution Pattern

Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.

Code Information

8018728 8080823 8314382 8409523 8023194 7857489