Z-0968-2013 Class I Terminated

Recalled by Del Mar Reynolds Medical, Ltd. — Hertford, N/A

FDA device recall Z-0968-2013 was initiated by Del Mar Reynolds Medical, Ltd. on Jan 15, 2013 and is designated Class I. Reason for recall: A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being… The recall status is terminated (terminated May 14, 2013). Affected quantity: 108 total (17 units in the US and 91 units outside the US).

Recall details

Product Type
Devices
Report Date
Mar 27, 2013
Initiation Date
Jan 15, 2013
Termination Date
May 14, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
108 total (17 units in the US and 91 units outside the US)

Product description

Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.

Reason for recall

A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.

Distribution pattern

Worldwide Distribution - USA Nationwide including the states of New Mexico, North Carolina and the countries of Argentina, China, Mexico, and Taiwan.

Code information

serial numbers: ABSO-004448, ABSO-004458, ABSO-004459, ABSO-004460, ABSO-004461, ABSO-004462, ABSO-004463, ABSO-004465, ABSO-004466, ABSO-004467, ABSO-004468, ABSO-004469, ABSO-004470, ABSO-004471, ABSO-004472, ABSO-004473, ABSO-004474, ABSO-004475, ABSO-004476, ABSO-004477, ABSO-004478, ABSO-004479, ABSO-004480, ABSO-004481, ABSO-004482, ABSO-004483, ABSO-004487, ABSO-004488, ABSO-004489, ABSO-004490, ABSO-004491, ABSO-004492, ABSO-004493, ABSO-004494, ABSO-004495, ABSO-004497, ABSO-004499, ABSO-004500, ABSO-004503, ABSO-004504, ABSO-004505, ABSO-004506, ABSO-004507, ABSO-004508, ABSO-004510, ABSO-004511, ABSO-004512, ABSO-004513, ABSO-004526, ABSO-004528, ABSO-004529, ABSO-004530, ABSO-004531, ABSO-004532, ABSO-004533, ABSO-004534, ABSO-004535, ABSO-004536, ABSO-004537, ABSO-004538, ABSO-004539, ABSO-004540, ABSO-004541, ABSO-004542, ABSO-004543, ABSO-004544, ABSO-004545, ABSO-004564, ABSO-004566, ABSO-004567, ABSO-004568, ABSO-004569, ABSO-005883, ABSO-005884, ABSO-005885, ABSO-005886, ABSO-005887, ABSO-005945, ABSO-005947, ABSO-005948, ABSO-005949, ABSO-005950, ABSO-100023, ABSO-100024, ABSO-100038, ABSO-100039, ABSO-100040, ABSO-100041, ABSO-100042, ABSO-100043, ABSO-100070, ABSO-100071, ABSO-100072, ABSO-100073, ABSO-100074, ABSO-100075, ABSO-100076, ABSO-100077, ABSO-100078, ABSO-100079, ABSO-100080, ABSO-100081, ABSO-100082, ABSO-100083, ABSO-100084, ABSO-100085, ABSO-100086, and ABSO-100087. Firm added ABSO 004581 - which was inSpacelaabs inventory, in transit to a Field Service Engineer, at the time the recall was initiated. the nonconforming absorber has been replaced with a conforming part and this absober serial number was added to the US Spacelabs Recall tracking. Five incorrect serialnumbers were included in the original serial numbers provided to FDA. This was a result of a data entry error. An updated US and Rest of the World customer/Serial number was provided to FDA 3/25/2013. The correction to the Rest of the World report is as follows: Incorrect SN Reported to FDA Correct, Nonconforming SN ABSO 005883 ABSO-004514 ABSO 005884 ABSO-004515 ABSO 005885 ABSO-004516 ABSO 005886 ABSO-004517 ABSO 005887 ABSO-004518

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0968-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Del Mar Reynolds Medical, Ltd.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Del Mar Reynolds Medical, Ltd.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Del Mar Reynolds Medical, Ltd. →

Source: openFDA record for Z-0968-2013. Report a data issue: hello@regdatalab.com.