Z-0970-2020 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-0970-2020 was initiated by Cook Inc. on January 4, 2019 and is designated Class II. Reason for recall: The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip. The recall status is terminated (terminated May 29, 2020). Affected quantity: 13388 total.

Recall Details

Product Type
Devices
Report Date
February 12, 2020
Initiation Date
January 4, 2019
Termination Date
May 29, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13388 total

Product Description

Urethral Dilation Balloon Catheter with Open Tip , Global Product No. G17844

Reason for Recall

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Distribution Pattern

Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.

Code Information

8494465 8513227 8537103 8537104 8552050 8552051 8567256 8576963 8632766 8639366 8659206 8683459 8754089 NS8513228