Z-0978-2019 Class I Terminated

Recalled by O-Two Medical Technologies, Inc. — Brampton, Ontario

Recall Details

Product Type
Devices
Report Date
April 10, 2019
Initiation Date
February 12, 2019
Termination Date
June 29, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
632

Product Description

e700 Automatic Transport Ventilator, Model Number 01EVE700

Reason for Recall

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Distribution Pattern

The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.

Code Information

EV70001 to EV70658