Z-0990-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 13, 2019
- Initiation Date
- January 8, 2019
- Termination Date
- June 3, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4
Product Description
Maquet XS Flat Screen Monitor Holder, SAT XS32 12 SPE, Configuration Part Number ARD2SAT00130A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Reason for Recall
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Distribution Pattern
The products were distributed to the following US states: AZ and TX.
Code Information
Serial Numbers 500013 500014 500096 500097