Z-1008-2024 Class II Ongoing
FDA device recall Z-1008-2024 was initiated by Brasseler USA I Lp on Jan 11, 2024 and is designated Class II. Reason for recall: The device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece. The recall status is ongoing. Affected quantity: 440 burs.
Recall details
- Product Type
- Devices
- Report Date
- Feb 14, 2024
- Initiation Date
- Jan 11, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 440 burs
Product description
Reason for recall
Distribution pattern
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, IL, NY, OH, OR, PA, TN, TX, and WA. The country of Canada.