Z-1010-2018 Class II Terminated

Recalled by Fresenius Medical Care Renal Therapies Group, LLC — Waltham, MA

Recall Details

Product Type
Devices
Report Date
March 21, 2018
Initiation Date
September 29, 2017
Termination Date
August 9, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
598 cases

Product Description

NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason for Recall

The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.

Distribution Pattern

US Distribution to states of: CA, FL, KS, LA, and TX.

Code Information

Lot # 17KTAC076; UDI-00840861101696