Z-1016-2018 Class II Terminated
FDA device recall Z-1016-2018 was initiated by Zimmer Biomet, Inc. on Nov 29, 2017 and is designated Class II. Reason for recall: The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed… The recall status is terminated (terminated Apr 2, 2020). Affected quantity: 13,227 in total.
Recall details
- Product Type
- Devices
- Report Date
- Mar 21, 2018
- Initiation Date
- Nov 29, 2017
- Termination Date
- Apr 2, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13,227 in total
Product description
Reason for recall
Distribution pattern
Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1016-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Zimmer Biomet, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Zimmer Biomet, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.