Z-1024-2025 Class II Ongoing
FDA device recall Z-1024-2025 was initiated by Boston Scientific Corporation on Dec 20, 2024 and is designated Class II. Reason for recall: Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for… The recall status is ongoing. Affected quantity: 495 units US; 2 units(OUS).
Recall details
- Product Type
- Devices
- Report Date
- Feb 5, 2025
- Initiation Date
- Dec 20, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 495 units US; 2 units(OUS)
Product description
Reason for recall
Distribution pattern
Nationwide including Puerto Rico Foreign: To be provided
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1024-2025
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026