Z-1024-2026 Class II Ongoing
FDA device recall Z-1024-2026 was initiated by REPLIGEN CORPORATION on Dec 10, 2025 and is designated Class II. Reason for recall: Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can… The recall status is ongoing. Affected quantity: 33.
Recall details
- Product Type
- Devices
- Report Date
- Jan 21, 2026
- Initiation Date
- Dec 10, 2025
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 33
Product description
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Reason for recall
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Distribution pattern
U.S.
Code information
PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1024-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026