Z-1024-2026 Class II Ongoing

Recalled by REPLIGEN CORPORATION — Marlborough, MA

FDA device recall Z-1024-2026 was initiated by REPLIGEN CORPORATION on Dec 10, 2025 and is designated Class II. Reason for recall: Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can… The recall status is ongoing. Affected quantity: 33.

Recall details

Product Type
Devices
Report Date
Jan 21, 2026
Initiation Date
Dec 10, 2025
Voluntary/Mandated
FDA Mandated
Product Quantity
33

Product description

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Reason for recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Distribution pattern

U.S.

Code information

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1024-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1024-2026. Report a data issue: hello@regdatalab.com.