Z-1024-2026 Class II Ongoing

Recalled by REPLIGEN CORPORATION — Marlborough, MA

Recall Details

Product Type
Devices
Report Date
January 21, 2026
Initiation Date
December 10, 2025
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
33

Product Description

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Reason for Recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Distribution Pattern

U.S.

Code Information

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496