Z-1026-2016 Class II Terminated

Recalled by Ivoclar Vivadent, Inc. — Buffalo, NY

FDA device recall Z-1026-2016 was initiated by Ivoclar Vivadent, Inc. on Feb 12, 2016 and is designated Class II. Reason for recall: Complaints were received claiming the material failed to polymerize under the curing light. After investigation, it was determined that a wrong photo-initiator was used in the production batch U30755… The recall status is terminated (terminated May 8, 2017). Affected quantity: Domestic: 406 units.

Recall details

Product Type
Devices
Report Date
Mar 9, 2016
Initiation Date
Feb 12, 2016
Termination Date
May 8, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Domestic: 406 units

Product description

Helioseal Clear Refill 1 x 1.25 g, REF/Product Code/Art. No. 558521AN and Helioseal Clear Assortment 5 x 1.25 g, REF/Product Code/Art. No. 558520AN ---- COMMON/USUAL NAME: Pit & Fissure Sealant

Reason for recall

Complaints were received claiming the material failed to polymerize under the curing light. After investigation, it was determined that a wrong photo-initiator was used in the production batch U30755 of Helioseal Clear. This failure can lead to incomplete curing of the sealant, if mono-wave LED curing lights are used.

Distribution pattern

Nationwide distribution including NY, PA, FL, IN, IA, TX, and CA.

Code information

REF/Product Code/Art. No. 558521AN, Lot U29336 (exp. 10/5/2020); REF/Product Code/Art. No. 558520AN, Lots U33966, U35661, U50469 (exp. 10/5/2020)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1026-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ivoclar Vivadent, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ivoclar Vivadent, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ivoclar Vivadent, Inc. →

Source: openFDA record for Z-1026-2016. Report a data issue: hello@regdatalab.com.