Z-1027-2017 Class II Terminated

Recalled by Fresenius Medical Care Renal Therapies Group, LLC — Waltham, MA

FDA device recall Z-1027-2017 was initiated by Fresenius Medical Care Renal Therapies Group, LLC on Dec 21, 2016 and is designated Class II. Reason for recall: When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate… The recall status is terminated (terminated Mar 17, 2023). Affected quantity: 78481 (US) 2734 units (OUS).

Recall details

Product Type
Devices
Report Date
Jan 25, 2017
Initiation Date
Dec 21, 2016
Termination Date
Mar 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
78481 (US) 2734 units (OUS)

Product description

Fresenius 2008K Series : Hemodialysis System

Reason for recall

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Distribution pattern

Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Code information

All serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1027-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fresenius Medical Care Renal Therapies Group, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Fresenius Medical Care Renal Therapies Group, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Fresenius Medical Care Renal Therapies… →

Source: openFDA record for Z-1027-2017. Report a data issue: hello@regdatalab.com.