Z-1028-2013 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-1028-2013 was initiated by Roche Diagnostics Operations, Inc. on Mar 1, 2013 and is designated Class II. Reason for recall: Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting… The recall status is terminated (terminated Jun 4, 2014). Affected quantity: 189 units.

Recall Details

Product Type
Devices
Report Date
Apr 10, 2013
Initiation Date
Mar 1, 2013
Termination Date
Jun 4, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
189 units

Product Description

Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas¿ CT/NG 4800 System, for in vitro diagnostics.

Reason for Recall

Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.

Distribution Pattern

Nationwide Distribution including Puerto Rico and the states of MO, NE, CO, MT, NJ, FL, WA, TX, GA, NY, KS, NH, VA, HI, PA, MI, IA, CA, NC, AZ, UT, LA, IL, TN, ME, MN, KY, CT, OR, NV, AL and NM.

Code Information

Lot / SN : 477415,477416, 477898, and 477899

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1028-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.