Z-1030-2019 Class II Terminated

Recalled by CTL Medical Corporation — Addison, TX

FDA device recall Z-1030-2019 was initiated by CTL Medical Corporation on Jul 5, 2018 and is designated Class II. Reason for recall: the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the… The recall status is terminated (terminated Apr 28, 2021). Affected quantity: 133 total devices.

Recall details

Product Type
Devices
Report Date
Mar 27, 2019
Initiation Date
Jul 5, 2018
Termination Date
Apr 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
133 total devices

Product description

CTL Medical Hex Driver, 1/4" Square QC, 4.0 Ratcheting Ref 015.7050

Reason for recall

the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.

Distribution pattern

distributor and user level

Code information

Lot 7EQAB-D, 7ATAB-D

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1030-2019
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1030-2019. Report a data issue: hello@regdatalab.com.