Z-1032-2022 Class I Terminated

Recalled by Avanos Medical, Inc. — Alpharetta, GA

FDA device recall Z-1032-2022 was initiated by Avanos Medical, Inc. on Mar 21, 2022 and is designated Class I. Reason for recall: Product labeling is being modified. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol. The recall status is terminated (terminated Sep 9, 2024). Affected quantity: 721 devices.

Recall Details

Product Type
Devices
Report Date
May 25, 2022
Initiation Date
Mar 21, 2022
Termination Date
Sep 9, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
721 devices

Product Description

Avanos Cortrak 2 Enteral Access System (EAS)

Reason for Recall

Product labeling is being modified. STOP using the Anonymous Account Mode feature of the CORTRAK 2 device immediately. CONFIRM placement of the NG/NI tube per institution protocol.

Distribution Pattern

Worldwide Distribution. US Nationwide, Australia, Austria, Canada, Germany, Ireland, Kuwait, New Zealand, India, Japan, Malaysia, Netherlands, Panama, Singapore, Switzerland, and United Kingdom.

Code Information

1) Product Code 20-0950, CORTRAK 2 Enteral Access System (EAS), UDI: 00350770472010; all serial numbers 2) Product Code P20-0950, CORTRAK 2 Enteral Access System (EAS) - Loaner Unit, UDI: 00350770472065; all serial numbers 3) Product Code 20-0950, CORTRAK 2 Enteral Access System (EAS) - Halyard version; UDI: 10680651472011; all serial numbers 4) Product Code P20-0950, CORTRAK 2 Enteral Access System (EAS) - Loaner Unit - Halyard version; UDI: 10680651472066; all serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1032-2022
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.