Z-1034-2017 Class II Terminated

Recalled by Ultradent Products, Inc. — South Jordan, UT

FDA device recall Z-1034-2017 was initiated by Ultradent Products, Inc. on Dec 14, 2016 and is designated Class II. Reason for recall: Ultradent Products, Inc. announces a voluntary field action for the NanoTips 30g 20pk, 100pk because the cannula attached to the tip hub may become loose or detach from the blue base that connects to… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 143 packs of 100 tips; 104 packs of 20 tips.

Recall details

Product Type
Devices
Report Date
Jan 25, 2017
Initiation Date
Dec 14, 2016
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
143 packs of 100 tips; 104 packs of 20 tips

Product description

NanoTips 30g 20pk, 100pk PN 1240, 1426 Description: NanoTips are injection molded royal blue tips that contain stainless steel 30-gauge cannula to aid in the delivery of chemicals. The NanoTip base is made from a polypropylene copolymer. The stainless steel cannula is made from 304 SS, and is 30-gauge in diameter and 0.488 in length. The cannula is secured in the base with an epoxy adhesive. Indications: Tips, when used in conjunction with the range of available syringes and products, provide a technique that is designed to provide a faster, easier, and more precise delivery. They are designed and created in such a way to generate less product waste and provide optimum control/leverage. Tips are also used for moisture removal and for directing laser light beam. Each of the tips is part of delivery systems designed for dentist-supervised, intra-oral procedures. The NanoTIp is designed for use with dental products manufactured by Ultradent Products, Inc. such as: UltraSeal XT¿ plus, UltraSeal XT¿ hydro, Ultra-Etch¿, PermaFloIdeal for microscopic restorative dentistry

Reason for recall

Ultradent Products, Inc. announces a voluntary field action for the NanoTips 30g 20pk, 100pk because the cannula attached to the tip hub may become loose or detach from the blue base that connects to the syringe.

Distribution pattern

Worldwide Distribution - US (Nationwide) - Internationally to Belgium, Switzerland, Germany, Italy, South Korea, Netherlands, and Japan.

Code information

Lot Numbers: BD6KM, BCVPG, BCTBY, BD3K4, BCWVZ

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1034-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ultradent Products, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1034-2017. Report a data issue: hello@regdatalab.com.