Z-1044-2016 Class II Terminated

Recalled by ConvaTec, Inc — Greensboro, NC

FDA device recall Z-1044-2016 was initiated by ConvaTec, Inc on Nov 11, 2015 and is designated Class II. Reason for recall: The Nebulizer mask malfunctions in that there is reduced or no aerosol output. The malfunction occurs from the reduction in the diameter of a component part inside the nebulizer. The recall status is terminated (terminated Jul 18, 2016). Affected quantity: 3,605,700 pieces.

Recall details

Product Type
Devices
Report Date
Mar 9, 2016
Initiation Date
Nov 11, 2015
Termination Date
Jul 18, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,605,700 pieces

Product description

Nebulizer Masks Opti-Mist and Opti-Mist Clear Nebulizer masks are intended to spray liquids in aerosol form into gasses that are delivered directly to a patient.

Reason for recall

The Nebulizer mask malfunctions in that there is reduced or no aerosol output. The malfunction occurs from the reduction in the diameter of a component part inside the nebulizer.

Distribution pattern

Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, IA, KS, ME, MA, MN, MS, MO, NV, NH, NJ, NY, NC, OH, OR, PA, SC, TX VA, WA, WI, and countries of: Australia, Belgium, Burma, Canada, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Netherlands, Norway, Russia, Saudi Arabia, Sweden, Switzerland, United Kingdom & Vietnam.

Code information

lot/batch numbers 13-23, 13-24, 13-29, 13-30, 13-31, 13-32, 97750, 97774, 97777, 97870, 97871, 97873, 97874, 97880, 97988, 98030, 98031, 98033, 98034, 98043, 98048, 98058, 98059, 98099, 98103, 98330, 98338, 98380, 98491, 98520, 98546, 98556, 98570, 98571, 98578, 98595, 98635, 98683, 98694, 98721, 98722, 98730, 98733, 98738, 98777, 98779, 98782, 98840, 98841, 98843, 98844, 98853, 98856, 98859, 98867, 98868, 98869, 98870, 98871, 98969, 99078, 99079, 99199, 99206, 99207, 99227, 99230, 99231, 99232, 99297, 99397, 99398, 99447, 99450, 99451, 99466, 99492, 99539, 99577, 99578, 99591, 99606, 99628, 99630, 99631, 99642, 99661, 99673, 99674, 99708, 99709, 99735, 99845, 99852, 99853, 99854, 99919, 99920, 99928, 99932, 99997, 100029, 100065, 100066, 100074, 100077, 100101, 100102, 100114, 100119, 100155, 100167, 100177, 100180, 100205, 100206, 100207, 100238, 100254, 100401, 100402, 100424, 100483, 100506, 100516, 100546, 100549, 100552, 100563, 100599, 100603, 100693, 100697, 100700, 100705, 100752, 100759, 100784, 100809, 100812, 100814, 100871, 100931, 100984, 100985, 101186, 101200, 101201, 101202, 101310, 101312, 101450, 101473, 101474, 101484, 101485, 101527, 101528, 101533, 101537, 101555, 101556, 101557, 101558, 101559, 101625, 101626, 101686, 101704, 101708, 101709, 101721, 101723, 101750, 101751, 101752, 101753, 101759, 101761, 101785, 101889, 109211, 101926, 101949, 101984, 102038, 102046, 102061, 102063, 102076, 102135, 102138, 102141, 102281, 102284, 102330, 102389, 102461, 102468, 102473, 102474, 102476, 102478, 102570, 102571, 102606, 102642, 102671, 102681, 102702, 102703, 102710, 102712, 102720, 102725, 102744, 102745, 102746, 102751, 102796, 102806, 102807, 102861, 102945, 102976, 103021, 103076, 103079, 103116, 103151, 103161, 103166, 103206, 103284, 103301, 103341, 103342, 103348, 103387, 103429, 103431, 103432, 103448, 103449, 103463, 103464, 103465, 103485, 103575, 103580, 103585, 103587, 103607, 103674, 103675, 103693, 103694, 103695, 103746, 103747, 103786, 103790, 103802, 103803, 103825, 103827, 103829, 103857, 103890, 103892, 103917, 103928, 103979, 103980, 104005, 104011, 104043, 104046, 104055, 104067, 104068, 104092, 104093, 104095, 104096, 104100, 104101, 104110, 104131, 104211, 104220, 104222, 104223, 104224, 104225, 104243, 104244, 104245, 104254, 104274, 104333, 104335, 104415, 104442, 104449, 104451, 104464, 104484, 104503, 104509, 104510, 104514, 104598, 194601, 104628, 104647, 104750, 104753, 104756, 104783, 104818, 104820, 104830, 104831, 104867, 104916, 104945, 105002, 105003, 105012, 105013, 105014, 105015, 105016, 105017, 105018, 105021, 105022, 105023, 105025, 105026, 105137, 105179, 105195, 105238, 105242, 105243, 105318, 105327, 105328, 105334, 105361, 105362, 105363, 105397, 105414, 105415, 105420, 105423, 105452, 105467, 105509, 105513, 105573, 105593, 105680, 105693, 105699, 105702, 105703, 105704, 105728, 105729, 105730, 105731, 105736, 105767, 105772, 105795, 105827, 105833, 105835, 105836, 105849, 105852, 105853, 105906, 105960, 105961, 105980, 105989, 106022, 106023, 106024, 106114, 106153, 106154, 106155, 106156, 106174, 106189, 106250, 106263, 106275, 105286, 106286, 106308, 106337, 106338, 106347, 106357, 106381, 106382, 106385, 106386, 106436, 106456, 106490, 106514, 106521, 106523, 106532, 106596, 106571, 106612, 106657, 106658, 106744, 106745, 106755, 106766, 106782, 106785, 106806, 106842, 106977, 106978, 106982, 106998, 107123, 107124, 107147, 107174, 107220, 107223, 107304, 107309, 107372, 107401, 107468, 107502, 107516, 107544, 107592, 107610, 107631, 107657, 107717, 107739, 107751, 107757, 107826, 107844, 107845, 107872, 107886, 107896, 107969, 107972, 107973, 107974, 108057, 108144, 108223, 108225, 108421, 108422, 108423, 108424, 108425, 108427, 108432, 108451, 108453, 108460, 108461, 108462, 108469, 108470, 108471, 108472, 108473, 108476, 108477, 108478, 108501, 108502, 108503, 108550, 108551, 108600, 108601, 108615, 108653, 108655, 108662, 108663, 108677, 108709, 108712, 108714, 108721, 108725, 108726, 108730, 108734, 108741, 108747, 108815, 108820, 108892, 108893, 108894, 108906, 108908, 108920, 108926, 108928, 108940, 108946, 108951, 108956, 108966, 108968, 108970, 108982, 108988, 109058, 109082, 109092, 109096, 109113, 109130, 109162, 109205, 109206, 109211, 109420, 109222, 109223, 109225, 109226, 109227, 109245, 109301, 109345, 109377, 109403, 109417, 109421, 109437, 109462, 109552, 109555, 109559, 109581, 109690, 109694, 109718, 109719, 109739, 109746, 109763, 109805, 109885, 109969, 109972, 109985, 109994, 110026, 110329, 110557, 111011, 111041, 111042, 111468, 111469, 111470, 111971, 112005, 112062

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1044-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ConvaTec, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1044-2016. Report a data issue: hello@regdatalab.com.