Z-1046-2013 Class III Terminated
FDA device recall Z-1046-2013 was initiated by Siemens Healthcare Diagnostics, Inc. on Feb 27, 2013 and is designated Class III. Reason for recall: Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation… The recall status is terminated (terminated Jan 14, 2014). Affected quantity: 2207 total.
Recall details
- Product Type
- Devices
- Report Date
- Apr 10, 2013
- Initiation Date
- Feb 27, 2013
- Termination Date
- Jan 14, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2207 total
Product description
Siemens Dimension(R) EXL(TM) LM (Siemens Material Number 10486890). Intended to measure a variety of analytes in human body fluids.
Reason for recall
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
Distribution pattern
Worldwide Distribution.
Code information
Siemens Material Number 10486890
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1046-2013
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026