Z-1047-2013 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-1047-2013 was initiated by Siemens Healthcare Diagnostics, Inc. on Feb 27, 2013 and is designated Class III. Reason for recall: Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation… The recall status is terminated (terminated Jan 14, 2014). Affected quantity: 2207 total.

Recall details

Product Type
Devices
Report Date
Apr 10, 2013
Initiation Date
Feb 27, 2013
Termination Date
Jan 14, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2207 total

Product description

Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.

Reason for recall

Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h

Distribution pattern

Worldwide Distribution.

Code information

Siemens Material Number 106393377

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1047-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1047-2013. Report a data issue: hello@regdatalab.com.