Z-1052-2013 Class I Terminated
FDA device recall Z-1052-2013 was initiated by Stryker Howmedica Osteonics Corp. on Nov 8, 2012 and is designated Class I. Reason for recall: With regards to ShapeMatch Cutting Guides using MRI Imaging, Stryker Orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing… The recall status is terminated (terminated Jan 11, 2021). Affected quantity: 7868 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 24, 2013
- Initiation Date
- Nov 8, 2012
- Termination Date
- Jan 11, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7868 units
Product description
ShapeMatch Cutting Guides Product Usage: The ShapeMatch Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee arthroplasty components intraoperatively and in guiding the marking of bone before cutting.
Reason for recall
With regards to ShapeMatch Cutting Guides using MRI Imaging, Stryker Orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing ShapeMatch Cutting Guides.
Distribution pattern
USA Nationwide Distribution
Code information
Catalog No: TR7100-L and TR7100-R Lot Code: All
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1052-2013
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026