Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-1066-2015 was initiated by Stryker Instruments Div. Of Stryker Corporation on Jan 7, 2015 and is designated Class II. Reason for recall: Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker. The recall status is terminated (terminated Mar 9, 2016). Affected quantity: 399.

Recall details

Product Type
Devices
Report Date
Feb 18, 2015
Initiation Date
Jan 7, 2015
Termination Date
Mar 9, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
399

Product description

System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.

Reason for recall

Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.

Distribution pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France.

Code information

Part No. 6126-120-000, Lots 13205, 13209, 13210, 13212

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1066-2015
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1066-2015. Report a data issue: hello@regdatalab.com.