Z-1073-2024 Class II Ongoing

Recalled by Philips North America — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
February 21, 2024
Initiation Date
December 15, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units: 16 units US; 4 units OUS

Product Description

ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref:706150

Reason for Recall

Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

Distribution Pattern

Nationwide Foreign: Australia Germany India Italy Malaysia Panama Thailand

Code Information