Z-1076-2015 Class II Terminated

Recalled by MAST Biosurgery USA, Inc — San Diego, CA

FDA device recall Z-1076-2015 was initiated by MAST Biosurgery USA, Inc on August 26, 2014 and is designated Class II. Reason for recall: The expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration date was August of 2017, while the bar code date read as July… The recall status is terminated (terminated February 24, 2015). Affected quantity: 49 pieces.

Recall Details

Product Type
Devices
Report Date
February 18, 2015
Initiation Date
August 26, 2014
Termination Date
February 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
49 pieces

Product Description

The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm (01) 18964050255027 (17) 170731(10)52804 LOT 52804 2017-08 REF25502-01 Product is wrapped in a foil pouch that is then sealed into an additional envelope.

Reason for Recall

The expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration date was August of 2017, while the bar code date read as July 31, 2017. The correct expiration date was July 31, 2017.

Distribution Pattern

Worldwide Distribution to the country of China.

Code Information

LOT 52804 2017-08