Z-1077-2014 Class II Terminated

Recalled by DePuy Orthopaedics, Inc. — Warsaw, IN

FDA device recall Z-1077-2014 was initiated by DePuy Orthopaedics, Inc. on Dec 9, 2013 and is designated Class II. Reason for recall: Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This… The recall status is terminated (terminated Jun 10, 2014). Affected quantity: 2.

Recall Details

Product Type
Devices
Report Date
Mar 5, 2014
Initiation Date
Dec 9, 2013
Termination Date
Jun 10, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product Description

PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.

Reason for Recall

Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.

Distribution Pattern

Worldwide Distribution-USA including the states of IA, CA, KY, and FL, and the countries of Finland, Germany, Sweden, and Czech Republic.

Code Information

Catalog Number 960015 Lot Numbers 7810268 and 7806929 (US) Lot Number 7806934 (OUS)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1077-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.