Z-1081-2016 Class II Terminated
FDA device recall Z-1081-2016 was initiated by Stryker Medical Division Of Stryker Corporation on Dec 3, 2015 and is designated Class II. Reason for recall: Potential for power cord to melt; hazards include electric shock and fire. The recall status is terminated (terminated May 24, 2017). Affected quantity: 2694.
Recall details
- Product Type
- Devices
- Report Date
- Mar 16, 2016
- Initiation Date
- Dec 3, 2015
- Termination Date
- May 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2694
Product description
Stryker Position Pro Mattress. Mattress, Air Flotation, Alternating Pressure. PositionPRO is a powered pressure relief support surface with low air loss (LAL) intended for medical purposes. PositionPRO consists of multiple air cells filled and emptied by an integrated control unit to provide changes in the distribution of body weight for pressure relief. PositionPRO offers turn assist. PositionPRO is for use with GoBed II¿, Secure II/3¿, InTouch¿, and Epic II¿ 84 in. x 35 in. flat deck frames. This surface is powered via an AC power cord that comes in either a 4 foot long option (PN QDF8087) or 15 foot long option (PN QDF8088). The 4 foot long option is used when the surface is plugged into the auxiliary outlet of the bed. The 15 foot long option is utilized when the surface needs to be plugged into a wall outlet. This product uses a power cord, manufactured by ElectriCord that plugs into the wall outlet or the auxiliary outlet of the bed that the surface is placed onto.
Reason for recall
Potential for power cord to melt; hazards include electric shock and fire.
Distribution pattern
Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Chile, France, Hong Kong, Italy, Mexico, South Korea, Switzerland, United Kingdom.
Code information
model number: 2920-000-000 units serialized