Z-1083-2024 Class II Ongoing

Recalled by SEDECAL SA — Algete, N/A

FDA device recall Z-1083-2024 was initiated by SEDECAL SA on Oct 14, 2021 and is designated Class II. Reason for recall: While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of… The recall status is ongoing. Affected quantity: 851 devices.

Recall details

Product Type
Devices
Report Date
Feb 21, 2024
Initiation Date
Oct 14, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
851 devices

Product description

MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007

Reason for recall

While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.

Distribution pattern

US Nationwide distribution. There was government distribution but no military distribution.

Code information

Model 712001 - UDI-DI 00884838028586, all serial numbers; Model 712002 - UDI-DI 00884838082373, all serial numbers; Model 712004 - UDI-DI 08436046002173, all serial numbers; Model 712006 - UDI-DI 00884838091504, serial numbers: SN20410556 and SN21410184; Model 712007 - UDI-DI)08436046002357, serial numbers: 26, 27, 19410038, 19410044, 19410046, 21410261, 21410266, 21410280, 21410288, 21410290, 21410291, 21410293, 21410294, 21410339, 21410340, 21410341, 21410342, 21410345, 22410022, 22410023, SN0410157, SN1410049, SN19410034, SN19410035, SN19410036, SN19410037, SN19410071, SN19410092, SN19410093, SN19410094, SN19410097, SN19410098, SN19410099, SN19410100, SN19410101, SN19410102, SN19410115, SN19410116, SN19410123, SN19410127, SN19410129, SN19410135, SN19410136, SN19410137, SN20410029, SN20410030, SN20410032, SN20410033, SN20410034, SN20410044, SN20410045, SN20410049, SN20410060, SN20410063, SN20410064, SN20410065, SN20410066, SN20410071, SN20410072, SN20410073, SN20410074, SN20410081, SN20410082, SN20410083, SN20410084, SN20410093, SN20410094, SN20410095, SN20410096, SN20410097, SN20410115, SN20410116, SN20410117, SN20410118, SN20410119, SN20410120, SN20410154, SN20410171, SN20410172, SN20410202, SN20410218, SN20410219, SN20410242, SN20410260, SN20410261, SN20410279, SN20410280, SN20410288, SN20410289, SN20410290, SN20410291, SN20410298, SN20410299, SN20410300, SN20410313, SN20410321, SN20410334, SN20410360, SN20410367, SN20410373, SN20410395, SN20410396, SN20410428, SN20410472, SN20410473, SN20410482, SN20410555, SN20410556, SN20410557, SN20410566, SN20410567, SN20410570, SN21410017, SN21410034, SN21410035, SN21410048, SN21410060, SN21410061, SN21410062, SN21410081, SN21410093, SN21410094, SN21410094, SN21410099, SN21410103, SN21410104, SN21410105, SN21410117, SN21410118, SN21410119, SN21410130, SN21410131, SN21410132, SN21410133, SN21410149, SN21410150, SN21410153, SN21410154, SN21410155, SN21410169, SN21410170, SN21410171, SN21410172, SN21410187, SN21410188, SN21410189, SN21410190, SN21410200, SN21410201, SN21410203, SN21410204, SN21410205, SN21410206, SN21410207, SN21410208, SN21410212, SN21410213, SN21410214, SN21410232, SNZY20410009, SNZY20410011, SNZY20410014, and SNZY204110015.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1083-2024
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for SEDECAL SA? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for SEDECAL SA: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about SEDECAL SA →

Source: openFDA record for Z-1083-2024. Report a data issue: hello@regdatalab.com.