Z-1085-2026 Class II Ongoing
FDA device recall Z-1085-2026 was initiated by Medline Industries, LP on Dec 16, 2025 and is designated Class II. Reason for recall: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the… The recall status is ongoing. Affected quantity: 96 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 21, 2026
- Initiation Date
- Dec 16, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 96 units
Product description
Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
Reason for recall
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Distribution pattern
US Nationwide distribution.
Code information
1) UDI/DI 10198459269196 (EA) 40198459269197 (CS), Lot Number 25EBD732; 2) UDI/DI 10198459326318 (EA) 40198459326319 (CS), Lot Number 25FBA103.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1085-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab