Z-1090-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-1090-2026 was initiated by Medline Industries, LP on Dec 16, 2025 and is designated Class II. Reason for recall: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the… The recall status is ongoing. Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Jan 21, 2026
Initiation Date
Dec 16, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Reason for recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Distribution pattern

US Nationwide distribution.

Code information

UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1090-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1090-2026. Report a data issue: hello@regdatalab.com.