Z-1095-2017 Class II Terminated

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-1095-2017 was initiated by Baxter Healthcare Corporation on Jan 4, 2017 and is designated Class II. Reason for recall: Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components… The recall status is terminated (terminated Mar 8, 2019). Affected quantity: 256 kits.

Recall details

Product Type
Devices
Report Date
Feb 1, 2017
Initiation Date
Jan 4, 2017
Termination Date
Mar 8, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
256 kits

Product description

Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to provide high flux and low flux hemodialysis, hemofiltration and ultrafiltration on patients weighing 15 Kilograms or more. The Phoenix system is to be used with either high or low permeability dialyzers. The device is intended to be used by trained operators when prescribed by a physician, in a chronic care dialysis facility or acute care unit

Reason for recall

Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.

Distribution pattern

Nationwide Distribution

Code information

Product code: 6997274; Lot codes: 160718E, 160720E, 160825E

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1095-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1095-2017. Report a data issue: hello@regdatalab.com.