Z-1104-2017 Class II Terminated

Recalled by Mesa Laboratories Inc — Omaha, NE

FDA device recall Z-1104-2017 was initiated by Mesa Laboratories Inc on Apr 4, 2016 and is designated Class II. Reason for recall: Product resistance has fallen outside of label claim within expiry. The recall status is terminated (terminated Jan 26, 2017). Affected quantity: 328/ 50-count/ boxes; 16,400 individual units.

Recall details

Product Type
Devices
Report Date
Feb 1, 2017
Initiation Date
Apr 4, 2016
Termination Date
Jan 26, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
328/ 50-count/ boxes; 16,400 individual units

Product description

ProSpore 4 mL Ampoule Biological indicator for steam sterilization, contains Geobacillus stearothermophilus, Catalog Number: PS-6-50, contains 50 Ampoules/Box. Product Usage: Biological indicator for steam sterilization at 121 deg C intended for use by a health care provider to accompany products being sterilized through steam sterilization procedure and to monitor adequacy of sterilization.

Reason for recall

Product resistance has fallen outside of label claim within expiry.

Distribution pattern

Product was distributed to: AL, AR, CA, FL, IA, MO, NE, NC, NY, OH, and WA. The product was also distributed internationally, including Mexico and Canada. There was no goverment or military distribution.

Code information

Lot # 637, exp. date 12/2016

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1104-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Mesa Laboratories Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1104-2017. Report a data issue: hello@regdatalab.com.