Z-1110-2020 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

Recall Details

Product Type
Devices
Report Date
February 19, 2020
Initiation Date
January 10, 2020
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16

Product Description

Terumo Advanced Perfusion Assembly Pump Guts, 6 Inch, Catalog Number 801798 - is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical proffessional who is experienced in the operation of this or similar equipment. The roller pump is used to move fluids through the cardiopulmonary bypass circuit and can be used for a number of applicatinos including arterial blood pumping, cardioplegia delivery, suction, and venting.

Reason for Recall

These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of CT, FL, LA, MA, MI, NY, and TX. Countries of Hong Kong, India, South Korea, and Taiwan.

Code Information

Serial Numbers: 882298 882299 882300 882752 882753 882794 882795 882796 882797 882798 882799 885547 885548 885549 885550 885553