Z-1113-2013 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-1113-2013 was initiated by Roche Diagnostics Operations, Inc. on Mar 6, 2013 and is designated Class II. Reason for recall: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if… The recall status is terminated (terminated Jan 10, 2018). Affected quantity: Both devices 4467.

Recall Details

Product Type
Devices
Report Date
Apr 24, 2013
Initiation Date
Mar 6, 2013
Termination Date
Jan 10, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Both devices 4467

Product Description

ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

Reason for Recall

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.

Distribution Pattern

Nationwide Distribution.

Code Information

part number 05060311001, 05060303001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1113-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.