Z-1119-2013 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

FDA device recall Z-1119-2013 was initiated by Siemens Healthcare Diagnostics, Inc on Mar 13, 2013 and is designated Class II. Reason for recall: Negative bias in patient samples for Vitamin D The recall status is terminated (terminated Mar 5, 2015). Affected quantity: 3871.

Recall details

Product Type
Devices
Report Date
Apr 24, 2013
Initiation Date
Mar 13, 2013
Termination Date
Mar 5, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3871

Product description

ADVIA Centaur Vitamin D Assay (100 tests); Reference Number: 10491994. For use in the quantitative determination of Vit D.

Reason for recall

Negative bias in patient samples for Vitamin D

Distribution pattern

Nationwide Distribution.

Code information

Ready Pack Lot Number: 134009 Kit Lot Number/Exp. Date: 88766009 4/28/2013 88786009 4/28/2013 89198009 4/28/2013 Ready Pack Lot Number: 134010 Kit Lot Number/Exp. Date: 90671010 6/2/2013 92212010 6/2/2013 Ready Pack Lot Number: 134011 Kit Lot Number/Exp. Date: 92395011 8/12/2013 94660011 8/12/2013 95016011 8/12/2013 95313011 8/12/2013

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1119-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc →

Source: openFDA record for Z-1119-2013. Report a data issue: hello@regdatalab.com.