Z-1122-2020 Class II Ongoing

Recalled by Flowonix Medical Inc — Mansfield, MA

FDA device recall Z-1122-2020 was initiated by Flowonix Medical Inc on Dec 24, 2019 and is designated Class II. Reason for recall: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of… The recall status is ongoing. Affected quantity: 6670 (US).

Recall details

Product Type
Devices
Report Date
Feb 19, 2020
Initiation Date
Dec 24, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6670 (US)

Product description

Prometra II 40mL Pump. Catalog No. 16827 The device is an implantable infusion pump. GTIN-14 00810335020242

Reason for recall

A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.

Distribution pattern

Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.

Code information

All units with pump firmware version 0.26 manufactured from 2012 until the present time. Serial Numbers 10GC4A24X 10GC4A02X 10GC4A18X 10GC4A22X 10GC4A14X 4MA00044X 4MA00006X 10GC4A12X 10GC4A09X 10GC4A06X 10GC4A21X 4MA00008X 10GC4A01X 10GC4A20X 10GC4A11X 10GC4A16X 10GC4A17X 4MA00045X 4MA00055X 10GC4A08X 4MA00032X 4MA00031X 4MA00010X 4MA00042X

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1122-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1122-2020. Report a data issue: hello@regdatalab.com.