Z-1131-2016 Class II Terminated

Recalled by Zimmer Manufacturing B.V. — Mercedita, PR

FDA device recall Z-1131-2016 was initiated by Zimmer Manufacturing B.V. on Jan 11, 2016 and is designated Class II. Reason for recall: As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. The recall status is terminated (terminated Apr 13, 2017). Affected quantity: 2236.

Recall details

Product Type
Devices
Report Date
Mar 23, 2016
Initiation Date
Jan 11, 2016
Termination Date
Apr 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2236

Product description

Product 33 consists of all product under product code: JDI and same usage: Item no: 662406515 HGP II ACETABULAR CUP BON 662406520 HGP II ACETABULAR CUP BON 662406525 HGP II ACETABULAR CUP BON 662406530 HGP II ACETABULAR CUP BON 662406535 HGP II ACETABULAR CUP BON 662406540 HGP II ACETABULAR CUP BON 662406550 HGP II ACETABULAR CUP BON 662406560 HGP II ACETABULAR CUP BON For use in total hip arthroplasty

Reason for recall

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Distribution pattern

US Nationwide Distribution

Code information

lot no.: 61674472 61906696 62016799 62332483 61674473 61902194 61906697 62326386 61674476 61953546 62161131 62172656 62326385 61953547 62011539 62017894 62130757 61886072 61953548 62017895 62172659 61886073 61948763 61953551 62172661 61874793 61948764 61996232 62017897 62172663 61874794 61953554 62286122

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1131-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Zimmer Manufacturing B.V.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1131-2016. Report a data issue: hello@regdatalab.com.