Z-1132-2013 Class II Terminated

Recalled by Smith & Nephew Inc — Memphis, TN

FDA device recall Z-1132-2013 was initiated by Smith & Nephew Inc on Mar 6, 2013 and is designated Class II. Reason for recall: One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter). The recall status is terminated (terminated Aug 16, 2019). Affected quantity: 23 units.

Recall details

Product Type
Devices
Report Date
Apr 24, 2013
Initiation Date
Mar 6, 2013
Termination Date
Aug 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23 units

Product description

TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery

Reason for recall

One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter).

Distribution pattern

Nationwide Distribution including CA, FL, GA, MA, MI, NC, SC, OH, and MO.

Code information

Lot Number 12FM12182

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1132-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Smith & Nephew Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Smith & Nephew Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Smith & Nephew Inc →

Source: openFDA record for Z-1132-2013. Report a data issue: hello@regdatalab.com.