Z-1136-2016 Class II Terminated
Recalled by Zimmer Manufacturing B.V.
— Mercedita, PR
FDA device recall Z-1136-2016 was initiated by Zimmer Manufacturing B.V. on Jan 11, 2016 and is designated Class II. Reason for recall: As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. The recall status is terminated (terminated Apr 13, 2017). Affected quantity: 4752.
Recall details
- Product Type
- Devices
- Report Date
- Mar 23, 2016
- Initiation Date
- Jan 11, 2016
- Termination Date
- Apr 13, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4752
Product description
Product 38 consists of all product under product code: HSB and same usage: Item no: 225303255 INTERLOCKING IM SCREW LG 225303555 INTERLOCKING IM SCREW LG 225303755 INTERLOCKING IM SCREW LG 225307055 INTERLOCKING IM SCREW LG 225302742 INTERLOCKING IM SCREW MED 225303242 INTERLOCKING IM SCREW MED 225307042 INTERLOCKING IM SCREW MED 225302737 INTERLOCKING IM SCREW SML 225303237 INTERLOCKING IM SCREW SML Product Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Distribution pattern
US Nationwide Distribution
Code information
lot no.: 61793328 62253390 61641813 62194666 61993625 62262085 62313056 62313056N 62021140 62312364 62312364N 61817404 61809837 61851878 61987145 62165485 62244841 62244854 61763549 61826386 61851889 61968098 62114111 62114112 62172486 61804389 61841502 61872521 61962448 61983689 62137121 62244852 61665045 61841504 62031683 62137150 62244853 61809838 62306323 62306323N 61830462 61809839 62189554 61804722 61830471 62165499 62165501 62189558 61809843
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1136-2016
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Zimmer Manufacturing B.V.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works