Z-1136-2018 Class II Terminated
Recalled by Smiths Medical ASD Inc.
— Minneapolis, MN
FDA device recall Z-1136-2018 was initiated by Smiths Medical ASD Inc. on Nov 13, 2017 and is designated Class II. Reason for recall: Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of… The recall status is terminated (terminated Sep 16, 2025). Affected quantity: 16,600 pumps total.
Recall details
- Product Type
- Devices
- Report Date
- Mar 28, 2018
- Initiation Date
- Nov 13, 2017
- Termination Date
- Sep 16, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16,600 pumps total
Product description
Medfusion Syringe Pump 3010 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.
Reason for recall
Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.
Distribution pattern
Worldwide distribution- US nationwide Distribution and to the countries of : Australia, Barbados, Canada, Hong Kong, Indonesia, Jamaica, New Zealand, Philippines, Saudi Arabia, Singapore, South Africa, Trinidad and Tobago, United Arab Emirates, and United Kingdom.
Code information
Product Codes 3010, 3010A with Model No. M03055, M02396, M02315, M02385, M02636, M02780, M02890, M03075, M03245, M03335, M03592, M03599, M03727, M03729, M03817, M03887, M04406, M04568, M04778, M04870, M05494, M05587, M05762, M05766, M05767, M05768, M05849, M06244, M07153, M07159, M07335, M07361, M07443, M07460, M08153, M08157, M08168, M08735, M08736, M08820, M08853, M08871, M08878, M09038, M09057, M09992, M10227, M10229, M10263, M10270, M10325, M10342, M10347, M10353, M10366, M10589, M10879, M10890, M10896, M10900, M11875, M11877, M11898, M11900, M11918, M11919
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1136-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Smiths Medical ASD Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Smiths Medical ASD Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Smiths Medical ASD Inc. →
Use it in ChatGPT
Get a free API key
How the connector works