Z-1139-2018 Class I Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-1139-2018 was initiated by Becton Dickinson & Company on Mar 22, 2018 and is designated Class I. Reason for recall: BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics… The recall status is terminated (terminated Jul 25, 2022). Affected quantity: 1,116,012,050.

Recall Details

Product Type
Devices
Report Date
Mar 28, 2018
Initiation Date
Mar 22, 2018
Termination Date
Jul 25, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,116,012,050

Product Description

BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI: 50382903678612); 9. 367862 (UDI: 50382903678629); 10. 367863 (UDI: 50382903678636); 11. 368021 (UDI: 50382903680219); 12. 368047 (UDI: 50382903680479); 13. 368054 (UDI: 50382903680547); 14. 368661 (UDI: 50382903686617); 15. 364300 (UDI: pre-UDI)

Reason for Recall

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

Distribution Pattern

Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.

Code Information

All lots. Catalog Number (Shelf Life): 1. 366401 (24 months); 2. 366450 (24 months); 3. 366643 (12 months); 4. 367835 (16 months); 5. 367841 (15 months); 6. 367844 (16 months); 7. 367856 (16 months); 8. 367861 (16 months); 9. 367862 (16 months); 10. 367863 (17 months); 11. 368021 (16 months); 12. 368047 (15 months); 13. 368054 (16 months); 14. 368661 (17 months); 15. 364300 (Requested)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1139-2018
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.