Z-1140-2022 Class II Terminated

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-1140-2022 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Oct 22, 2021 and is designated Class II. Reason for recall: Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be… The recall status is terminated (terminated Jan 4, 2023). Affected quantity: 6 boxes.

Recall details

Product Type
Devices
Report Date
Jun 1, 2022
Initiation Date
Oct 22, 2021
Termination Date
Jan 4, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 boxes

Product description

ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954

Reason for recall

Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.

Distribution pattern

US Distribution to North Carolina.

Code information

Model Number: HII1220001954; Lot Numbers: 215907578 (Expiration Date 09/30/2022), 215907729 (Expiration Date 02/28/2023); Manufacturer Number: A11A01954

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1140-2022
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for MEDLINE INDUSTRIES, LP - Northfield? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for MEDLINE INDUSTRIES, LP - Northfield: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about MEDLINE INDUSTRIES, LP - Northfield →

Source: openFDA record for Z-1140-2022. Report a data issue: hello@regdatalab.com.