Z-1141-2018 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 28, 2018
- Initiation Date
- March 22, 2018
- Termination Date
- July 25, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 64,784,900
Product Description
BD Vacutainer¿ EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)
Reason for Recall
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
Distribution Pattern
Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
Code Information
All lots. Catalog Number (Shelf Life): 1. 367842 (Requested); 2. 367899 (Requested); 3. 368049 (Requested); 4. 368589 (Requested)