Z-1141-2024 Class II Ongoing
FDA device recall Z-1141-2024 was initiated by HeartSine Technologies Ltd on Feb 14, 2024 and is designated Class II. Reason for recall: Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR… The recall status is ongoing. Affected quantity: 12.
Recall details
- Product Type
- Devices
- Report Date
- Feb 21, 2024
- Initiation Date
- Feb 14, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 12
Product description
Reason for recall
Distribution pattern
Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1141-2024
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for HeartSine Technologies Ltd? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.