Z-1143-2023 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — East Walpole, MA

Recall Details

Product Type
Devices
Report Date
March 1, 2023
Initiation Date
December 22, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
158 units

Product Description

ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection of antibodies to the e antigen of the hepatitis B virus (HBV) in human pediatric (2-21 years old) and adult serum, EDTA plasma, or lithium heparin. Siemens Material Number (SMN): 10720831

Reason for Recall

Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2 testing, regardless of whether both tests are performed on the same sample

Distribution Pattern

US Nationwide distribution in the states of GA, UT.

Code Information

UDI: (01)00630414167923(10)B34847(17)20230309 (01)00630414167923(10)B34848(17)20230309 (01)00630414167923(10)B35113(17)20230629. Lot Numbers: B34847, B34848, B35113. Affects all current and future lots of the ADVIA Centaur XP and ADVIA Centaur XPT HBeAg assay until a solution is Implemented.