Z-1153-2020 Class II Ongoing

Recalled by Respironics California, LLC — Carlsbad, CA

Recall Details

Product Type
Devices
Report Date
February 19, 2020
Initiation Date
September 11, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
56

Product Description

Respironics RP-UI Assy, V60 Ventilators User Interface RP-Kits, REF/P/N: 453561528931, 453561528941, 453561536451, 453561536461

Reason for Recall

Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.

Distribution Pattern

US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam

Code Information

Serial Numbers: 100270400, 100270407, 100256220, 100270404, 100263004, 100266507, 100270402, 100263001, 100263009, 100258193, 100258197, 100258883, 100262978, 100263006, 100258196, 100256196, 100256197, 100256198, 100263003, 100266499, 100266505, 100258889, 100256192, 100258887, 100266509, 100270408, 100262976, 100258881, 100258880, 100266511, 100270406, 100258885, 100256223, 100263000, 100258884, 100258888, 100266500, 100266504, 100258886, 100256222, 100256195, 100256199, 100266502, 100266508, 100266510, 100270403, 100270490, 100257708, 100257713, 100257716, 100256831, 100256834, 100256837, 100262373, 100262556, 100262575, 100265136