Z-1162-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 1, 2023
- Initiation Date
- December 27, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 720 trays
Product Description
(1) Medline Sterile Maximum Barrier Pack-LF, Item DYNJ0248546C, 6 kits per case, contains BD Chloraprep Hi-Lite Orange, 10.5 mL Applicator, Vendor PN: 930715NS; (2) Medline Sterile Maximum Barrier Pack-LF, Item DYNJ0248546D, 6 kits per case, contains BD Chloraprep Hi-Lite Orange, 10.5 mL Applicator, Vendor PN: 930715NS.
Reason for Recall
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
Distribution Pattern
US Nationwide distribution in the states of AR, CO, and GA. There was no foreign/military/government distribution.
Code Information
(1) UDI/DI (01) 40889942055518; Lot numbers 21KBE911, exp. 3/31/2023; 21JBU806, exp. 3/31/2023; 21IBW034, exp. 3/31/2023; 21HBM415, exp. 3/31/2023; 21GBH194, exp. 3/31/2023; 21CBY712, exp. 3/31/2023; 21BBS824, exp. 3/31/2023; 21BBS043, exp. 3/31/2023; 20XBC895, exp. 12/31/2022; 20WBB357, exp. 8/31/2022; 20KBD323, exp. 9/30/2022; and 20JBT517, exp. 12/31/2021. (2) UDI/DI (01) 40195327069286; Lot number 22CBJ633, exp. 12/31/2022.