Z-1167-2020 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 19, 2020
- Initiation Date
- September 11, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5897
Product Description
Ultrasound Transducer (M2736A), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips M2736A Avalon Smart Ultrasound Bottom Case, NFPHA9515
Reason for Recall
Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.
Code Information
All serial numbers serviced between 9/2/15 and 4/2/19.