Z-1168-2014 Class II Terminated

Recalled by Biomet 3I, LLC — Palm Beach Gardens, FL

FDA device recall Z-1168-2014 was initiated by Biomet 3I, LLC on Jul 22, 2013 and is designated Class II. Reason for recall: During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required. The recall status is terminated (terminated Apr 21, 2015). Affected quantity: 170.

Recall Details

Product Type
Devices
Report Date
Mar 19, 2014
Initiation Date
Jul 22, 2013
Termination Date
Apr 21, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
170

Product Description

Tapered Navigator Certain Surgical Kit Product Usage: Molded plastic tray used in holding various dental instruments for autoclave processing.

Reason for Recall

During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required.

Distribution Pattern

Worldwide Distribution - USA Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, including Hawaii and Puerto Rico and the countries of : Argentina, Austria, Canada, China, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain, Hong Kong, Hungary, Israel, Japan, Korea, Mexico, Poland, Portugal, Paraguay, Sweden, Singapore, Taiwan, and Uruguay.

Code Information

SGTIKIT (SGTTRAY)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1168-2014
openFDA file date
Oct 26, 2026
Loaded by RegDataLab