Z-1184-2016 Class II Terminated
FDA device recall Z-1184-2016 was initiated by Stryker Medical Division Of Stryker Corporation on Feb 5, 2016 and is designated Class II. Reason for recall: The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance… The recall status is terminated (terminated Sep 6, 2016). Affected quantity: 36.
Recall details
- Product Type
- Devices
- Report Date
- Mar 23, 2016
- Initiation Date
- Feb 5, 2016
- Termination Date
- Sep 6, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 36
Product description
Prime Series Stretchers (Fifth Wheel, Big Wheel and Zoom) Powered stretcher
Reason for recall
The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we
Distribution pattern
Distributed in DC and the states of AL, CA, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, and WV.
Code information
The hydraulic jack assemblies (Part numbers: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as service parts on Stretchers assembled from 8/28/15 - 8/29/15