Z-1184-2025 Class II Ongoing

Recalled by Physio-Control, Inc. — Redmond, WA

FDA device recall Z-1184-2025 was initiated by Physio-Control, Inc. on Jan 21, 2025 and is designated Class II. Reason for recall: Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system. The recall status is ongoing. Affected quantity: 622.

Recall details

Product Type
Devices
Report Date
Feb 26, 2025
Initiation Date
Jan 21, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
622

Product description

LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo¿ LNCS-II rainbow¿ DCI 8¿ SpCO, Adult Reusable Sensor REF 11996-000519 Masimo¿ LNCS-II rainbow¿ DCI 8¿ SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK¿ 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8¿ SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO¿), and methemoglobin saturation (SpMet¿) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

Reason for recall

Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.

Distribution pattern

U.S. Nationwide distribution including in the states of CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MS, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WV.

Code information

Catalog Number: 11996-000515 UDI-DI/GTIN code: 00843997010863 Lot Number: 23G79 Catalog Number: 11996-000519 UDI-DI/GTIN code: 00843997015608 Lot Numbers: 23HER 23HPT 23JBN 23JTA Catalog Number: 11996-000520 UDI-DI/GTIN code: 00843997015615 Lot Numbers: 23HNV 23JBP

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1184-2025